Product Search
 

ITL 900

< back

 

To the limits of current measurement accuracy

 


 

 

 

 

 

 

Main characteristics:

  • Nominal current measurement (IPN): 400 ARMS
  • Measuring range: +/- 900 A
  • Excellent linearity: 3 ppm in the operating temperature range
  • High resolution
  • Low Initial offset: 20 ppm
  • Very low offset drift less than 0.3 ppm/K
  • Overall accuracy @ IPN @ +25°C: < +/- 0.0087 %
  • Wide frequency bandwidth up to 200 kHz (+/- 3 dB)

Features:

  • DC, AC pulse currents' measurements with galvanic isolation
  • Fluxgate technology
  • Electrostatic shield between primary and secondary circuit
  • Power supply +/- 15 Volt
  • Operating temperature range from +10 to +50°C
  • D-Sub 9 male interface output
  • Current output
  • Really quick response time (< 500 ns)
  • Really low output noise

Standard compliance:

  • European Directive EMC 2004 / 108 / EEC: CE Marking
  • EN 50178 for industrial applications
  • Typical applications:
    • Feedback element in high performance gradient amplifiers for MRI
    • Feedback element in precision current regulated devices (power supplies…)
    • Precise and high stability inverters
    • Medical equipment
    • Energy measurement
    • Power analyzers
    • Calibration units

Remarks:

  • Current overload capability
  • Can be synchronized with an external clock signal
  • Additional output indicating the transducer state
  • Additional output indicating when the transducer is synchronized with an external signal
  • A LED shows normal operation

 

Product search results
Model Expand Primary
Nominal
Value
Measuring
Range
Secondary
Signal
Supply
Voltage (V)
Accuracy (%) Mounting
 
ITL 900-T expand 400 A 900 A 266 mA 15-15 Bi 0.0087 Panel
  Technology : IT